FDA Adverse Event Injury Summary report: N

TECNIS SIMPLICITY

MDR report key: 21971798 · Received May 6, 2025

Report

Report Number
3012236936-2025-000126
Event Type
Injury
Date Received
May 6, 2025
Date of Event
March 17, 2025
Report Date
June 20, 2025
Manufacturer
AMO MANUFACTURING NETHERLANDS
Product Code
MFK
UDI-DI
05050474810778
PMA / PMN Number
P980040
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3: THE DEVICE WAS NOT RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. A REVIEW OF RECORDS INCLUDING DEVICE HISTORY AND COMPLAINT TRENDING WILL BE PERFORMED. UPON COMPLETION OF THE REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION A SUPPLEMENTAL MEDWATCH WILL BE FILED. ATTEMPTS HAVE BEEN MADE TO OBTAIN MISSING INFORMATION; HOWEVER, TO DATE, NO RESPONSE HAS BEEN RECEIVED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: SECTION D9: DEVICE AVAILABLE FOR EVALUATION: YES. SECTION D9: RETURNED TO MANUFACTURER ON: 5/6/2025. SECTION H3: DEVICE EVALUATED BY MANUFACTURER: YES. DEVICE EVALUATION: VISUAL INSPECTION UNDER MAGNIFICATION WAS PERFORMED ON THE SUSPECT PRODUCT AND FOUND THE LENS WAS CUT IN HALF. NO ISSUES WERE OBSERVED ON EITHER HALF OF THE LENS. NO FURTHER EVALUATION WAS PERFORMED. CONCLUSION: NO ISSUES THAT COULD CAUSE OR CONTRIBUTE TO THE COMPLAINT ISSUES REPORTED WERE IDENTIFIED DURING PRODUCT EVALUATION. THE ¿LENS CUT¿ OBSERVED DURING THE PRODUCT EVALUATION COULD NOT BE CONFIRMED TO BE RELATED TO THE MANUFACTURING OR DESIGN PROCESS. BASED ON THE COMPLAINT INVESTIGATION RESULTS, THE PRODUCT WAS RELEASED WITHIN SPECIFICATIONS. NO PRODUCT DEFICIENCY OR PRODUCT MALFUNCTION COULD BE IDENTIFIED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PRELOADED INTRAOCULAR LENSES (IOL) WERE EXPLANTED FROM PATIENT'S LEFT AND RIGHT EYE AS THE PATIENT WAS UNABLE NEUROADAPT TO THE LENSES. THE LENSES WERE EXPLANTED AND REPLACED WITH A NON JNN LENSES IN A SECONDARY PROCEDURE. THERE WAS NO PATIENT INJURY. THERE WAS NO MEDICAL/SURGICAL INTERVENTION REQUIRED. NO OTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
84602 TECNIS SIMPLICITY LENS, MULTIFOCAL INTRAOCULAR MFK AMO MANUFACTURING NETHERLANDS DRN00V 05050474810778

Patients

Seq Age Sex Outcome Treatment
1 41 YR Female Other| R