TECNIS SIMPLICITY
Report
- Report Number
- 3012236936-2025-000116
- Event Type
- Malfunction
- Date Received
- May 5, 2025
- Date of Event
- April 3, 2025
- Report Date
- July 1, 2025
- Manufacturer
- AMO PUERTO RICO MFG. INC.
- Product Code
- HQL
- UDI-DI
- 05050474739451
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
. SECTION E1 - TELEPHONE NUMBER: (B)(6). SECTION G4: PMA/510(K) NUMBER: THIS REPORT IS BEING FILED ON AN INTERNATIONAL DEVICE; TECNIS OPTIBLUE SIMPLICITY PRELOADED 1-PIECE IOL, MODEL DIB00V THAT HAS A SIMILAR DEVICE, TECNIS SIMPLICITY PRELOADED 1-PIECE IOL MODEL DIB00 WHICH IS DISTRIBUTED IN THE UNITES STATES UNDER PMA P980040. THE INTRAOCULAR LENS (IOL) WAS NOT RETURNED FOR EVALUATION. THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORD AND COMPLAINT TRENDING FOR THIS DEVICE WILL BE PERFORMED. UPON COMPLETION OF THE REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
ADDITIONAL INFORMATION: SECTION D9: DEVICE AVAILABLE FOR EVALUATION? YES. SECTION D9: DATE RETURNED TO MANUFACTURER: JUNE 12, 2025. SECTION H3: DEVICE EVALUATED BY MANUFACTURER? YES. DEVICE EVALUATION: VISUAL INSPECTION UNDER MAGNIFICATION WAS PERFORMED ON THE SUSPECT PRODUCT. A BLACK FIBER-LIKE MATERIAL WAS FOUND INSIDE THE RECEIVED GAUZE. THE MATERIAL IS A PROTEIN-BASED FIBER CONSISTENT WITH HUMAN HAIR. THE FOURIER TRANSFORM INFRARED (FTIR) SPECTRUM RAW DATA OUTPUT FILE WAS COMPARED AGAINST THE MANUFACTURING PROCESS FTIR LIBRARY AND DID NOT YIELD ANY RESULTS WITH AT LEAST A 0.90000 CORRELATION. THE COMPLAINT ISSUE "FOREIGN MATERIAL: LOOSE" WAS IDENTIFIED DURING PRODUCT EVALUATION; HOWEVER, BASED ON THE COMPLAINT INVESTIGATION RESULTS THE COMPLAINT ISSUE COULD NOT BE CONFIRMED TO BE RELATED TO THE MANUFACTURING OR DESIGN PROCESS. THEREFORE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY OR PRODUCT MALFUNCTION. CONCLUSION: BASED ON THE COMPLAINT INVESTIGATION RESULTS, THE PRODUCT WAS RELEASED WITHIN SPECIFICATIONS. NO PRODUCT DEFICIENCY OR PRODUCT MALFUNCTION WERE FOUND. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IT WAS REPORTED THAT WHEN THE PRELOADED INTRAOCULAR LENS (IOL) WAS IMPLANTED INTO THE PATIENT'S LEFT EYE, A BLACK FIBROUS FOREIGN MATERIAL WAS ALSO INSERTED. THE FOREIGN MATERIAL WAS REMOVED, AND THE OPERATION WAS COMPLETED. NO PATIENT INJURY WAS REPORTED FOLLOWING THE PROCEDURE. NO MEDICAL INTERVENTION WAS REQUIRED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1832448 | TECNIS SIMPLICITY | INTRAOCULAR LENS | HQL | AMO PUERTO RICO MFG. INC. | DIB00V | 05050474739451 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Male |