FDA Adverse Event Malfunction Summary report: N

DRILLING GUIDE DEPTH GAUGE

MDR report key: 21963607 · Received May 5, 2025

Report

Report Number
3014128390-2025-00019
Event Type
Malfunction
Date Received
May 5, 2025
Date of Event
April 17, 2025
Report Date
May 5, 2025
Manufacturer
FX SHOULDER SOLUTIONS DALLAS
Product Code
LXH
UDI-DI
03701037316894
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CONTAMINANT MATTER WAS DISCOVERED INSIDE THE CANNULA OF AN INSTRUMENT IN THE FIELD, AS THE PRODUCT WAS NOT SATISFACTORILY INSPECTED AND DECONTAMINATED. CAPA-2025-0030 HAS BEEN OPENED TO ADDRESS THIS ISSUE.

Description of Event or Problem · 0

DURING A CASE WITH THE SURGEON ON (B)(6) 2025, THE DRILL GUIDE WAS DISCOVERED TO HAVE BONE DEBRIS OBSTRUCTING THE CANNULA. THERE WERE 20 MINUTES OF SURGICAL DELAY AS THE SURGICAL TEAM USED ANOTHER SET OF INSTRUMENTATION TO CONTINUE THE SURGERY. THE INSTRUMENT SET IN QUESTION HAS BEEN IDENTIFIED. IT WAS SHIPPED TO THE DISTRIBUTOR 14-APR-2025, AND RETURNED TO FX SHOULDER SOLUTIONS ON 23-APR-2025. THE CASE PERFORMED ON 17-APR-2025 WAS THE ONLY CASE PERFORMED BY THE DISTRIBUTOR. THE INSTRUMENT IN QUESTION WAS NOT ADEQUATELY INSPECTED OR CLEANED BY FX SHOULDER SOLUTIONS WAREHOUSE STAFF. CAPA-2025-0030 HAS BEEN OPENED TO ADDRESS THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
423860 DRILLING GUIDE DEPTH GAUGE ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH FX SHOULDER SOLUTIONS DALLAS T0505 03701037316894

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown