FDA Adverse Event Malfunction Summary report: N

ARCHITECT ANTI-HCV

MDR report key: 21963396 · Received May 5, 2025

Report

Report Number
3002809144-2025-00151
Event Type
Malfunction
Date Received
May 5, 2025
Date of Event
April 16, 2025
Report Date
June 3, 2025
Manufacturer
ABBOTT GMBH
Product Code
MZO
UDI-DI
00380740112622
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

COMPLETED INFORMATION FOR SECTION A: 58 YEAR OLD MALE PATIENT: SID (B)(6) 84 YEAR OLD FEMALE PATIENT: SID (B)(6). ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 6C37 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 1L79, WITH 510K/PMA/BLA NUMBER P050042.

Additional Manufacturer Narrative · 0

RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE WITHOUT INDICATION FOR ANY ADDITIONAL ISSUE. REVIEW OF TRACKING AND TRENDING FOR THE ARCHITECT ANTI-HCV ASSAY DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY RELATED TO THE COMPLAINT ISSUE. ADDITIONALLY, AN INCREASE IN COMPLAINT ACTIVITY WAS NOT IDENTIFIED FOR REAGENT LOT 69348BE00 AND COMPLAINT ISSUE. DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NONCONFORMANCES, POTENTIAL NONCONFORMANCES, OR DEVIATIONS ASSOCIATED WITH LOT NUMBER 69348BE00. IN-HOUSE TESTING OF A RETAINED REAGENT KIT OF LOT 69348BE00 WAS PERFORMED. ALL SPECIFICATIONS WERE MET, AND NO FALSE NON-REACTIVE RESULTS WERE OBTAINED, SHOWING THAT THE LOTS GENERATE THE EXPECTED RESULTS. IN ADDITION, THE CLINICAL SENSITIVITY OF THE LOT WAS EVALUATED BY TESTING ONE COMMERCIALLY AVAILABLE SEROCONVERSION PANELS (ZEPTOMETRIX HCV 9045). THE SEROCONVERSION PANEL RESULTS WERE COMPARED TO HISTORICAL ARCHITECT ANTI-HCV TEST RESULTS PROVIDED BY ZEPTOMETRIX. THE LOT DETECTED THE SAME BLEEDS AS REACTIVE FOR THE SEROCONVERSION PANELS. BASED ON THESE DATA IT WAS SHOWN THAT THE SENSITIVITY PERFORMANCE OF THE COMPLAINT LOT IS NOT ADVERSELY AFFECTED. LABELING WAS REVIEWED AND FOUND TO ADEQUATELY ADDRESS THE ISSUE. BASED ON OUR INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WITH THE ARCHITECT ANTI-HCV FOR LOT 69348BE00 WAS IDENTIFIED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSE NONREACTIVE ARCHITECT ANTI-HCV RESULTS FOR MULTIPLE PATIENTS. THE FOLLOWING INFORMATION WAS PROVIDED: 58 YEAR OLD MALE PATIENT: (B)(6) 2025, SID (B)(6): INITIAL RESULT = 0.35 S/CO, REPEAT = 2.18 S/CO, HISTORICAL RESULT FROM (B)(6) 2024 = 2.57 S/CO. 84 YEAR OLD FEMALE PATIENT: (B)(6) 2025, SID (B)(6): INITIAL RESULT = 0.25 S/CO, REPEAT = 1.82 S/CO NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSE NONREACTIVE ARCHITECT ANTI-HCV RESULTS FOR MULTIPLE PATIENTS. THE FOLLOWING INFORMATION WAS PROVIDED: 58 YEAR OLD MALE PATIENT: (B)(6) 2025, SID (B)(6): INITIAL RESULT = 0.35 S/CO, REPEAT = 2.18 S/CO, HISTORICAL RESULT FROM (B)(6) 2024 = 2.57 S/CO. 84-YEAR-OLD FEMALE PATIENT: (B)(6) 2025, SID (B)(6): INITIAL RESULT = 0.25 S/CO, REPEAT = 1.82 S/CO NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468711 ARCHITECT ANTI-HCV ANTI-HCV MZO ABBOTT GMBH 69348BE00 00380740112622

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ARC I2K PROC MOD, 03M74-01, ISR60316| ARC I2K PROC MOD, 03M74-01, ISR60316