FDA Adverse Event Injury Summary report: N

GLP LOADER MODULE

MDR report key: 21961052 · Received May 5, 2025

Report

Report Number
3023268435-2025-00006
Event Type
Injury
Date Received
May 5, 2025
Date of Event
April 15, 2025
Report Date
May 20, 2025
Manufacturer
ABBOTT AUTOMATION SOLUTIONS GMBH
Product Code
JQP
UDI-DI
00380740204143
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THIS REPORT IS BEING FILED ON THE GLP LOADER MODULE, LIST NUMBER 06T12-51, WHICH HAS A SAME/SIMILAR COMPONENT OF THE MODULAR GLP SYSTEMS TRACK REGISTERED IN THE US, LIST NUMBER 04Z96-51, 510K K213486.

Additional Manufacturer Narrative · 0

THE FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THAT DURING INSTALLATION OF THE GLP LOADER MODULE (LM), SERIAL (B)(6), WHILE TRYING TO STAND UP, THEY LOST THEIR BALANCE AND TRIED TO BRACE THEIR HAND WHEN IT CONTACTED A SHARP SUPPORT STRUCTURE. DUE TO THE POSITIONING OF THE FSR DURING THE INCIDENT, THEIR HAND STRUCK DOWNWARD AND THEN LATERAL ON THE METAL STRUCTURE AT THE BOTTOM OF THE MODULE, CREATING 3 LACERATIONS TO THE FINGERS. TWO OF THE FINGERS REQUIRED STITCHES AND IT WAS NOTED THAT THE THIRD WAS A SUPERFICIAL CUT. NO FURTHER TREATMENT BY THE FSR WAS REPORTED. THE FSR ALSO STATED THAT NO SAFETY GLOVES WERE WORN AT THE TIME OF THE INCIDENT. CURRENT LABELING PROVIDES WARNINGS AND CAUTIONS SERVICE PERSONNEL TO WEAR SAFETY GLOVES TO AVOID CUT INJURIES WHILE PERFORMING THE INSTALLATION OF THE GLP LOADER MODULE FOR THE GLP ARCHIVE II. A REVIEW OF COMPLAINT AND TRENDING DATA FOR THE GLP LOADER MODULE DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY FOR THE ISSUE DESCRIBED IN THIS COMPLAINT. A REVIEW OF THE MANUFACTURING DOCUMENTATION DID NOT IDENTIFY ANY ISSUES ASSOCIATED WITH THE COMPLAINT ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. THE ISSUE IS CONSIDERED A USE ERROR AS THE FIELD SERVICE PERSONNEL DID NOT WEAR SAFETY GLOVES DURING INSTALLATION OF THE GLP LOADER MODULE. BASED ON THE AVAILABLE INFORMATION THE GLP LOADER MODULE, SERIAL (B)(6), IS PERFORMING AS INTENDED. NO SYSTEMIC ISSUE OR DEFICIENCY OF THE GLP LOADER MODULE WAS IDENTIFIED.

Description of Event or Problem · 0

THE FIELD SERVICE REPRESENTATIVE (FSR) WAS AT THE CUSTOMER SITE FOR A NEW INSTALL OF THE GLP LOADER MODULE. THE FSR LOST HER BALANCE AND TRIED TO BRACE HERSELF. HER HAND HIT A SHARP STRUCTURE ON THE GLP LOADER MODULE AND CUT THREE OF HER FINGERS. TWO OF HER FINGERS REQUIRED STITCHES. NO FURTHER IMPACT OR TREATMENT TO THE FSR WAS REPORTED.

Description of Event or Problem · 0

THE FIELD SERVICE REPRESENTATIVE (FSR) WAS AT THE CUSTOMER SITE FOR A NEW INSTALL OF THE GLP LOADER MODULE. THE FSR LOST HER BALANCE AND TRIED TO BRACE HERSELF. HER HAND HIT A SHARP STRUCTURE ON THE GLP LOADER MODULE AND CUT THREE OF HER FINGERS. TWO OF HER FINGERS REQUIRED STITCHES. NO FURTHER IMPACT OR TREATMENT TO THE FSR WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1820045 GLP LOADER MODULE CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE JQP ABBOTT AUTOMATION SOLUTIONS GMBH 00380740204143

Patients

Seq Age Sex Outcome Treatment
1 NA Female Required Intervention