FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2196059 · Received August 9, 2011

Report

Report Number
2531779-2011-05688
Event Type
Malfunction
Date Received
August 9, 2011
Report Date
May 12, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2011. (B)(6). CORRECTION NUMBER: 2531779-03/24/2010-003-R. DURING EVALUATION THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE PHYSICALLY DAMAGED.

Description of Event or Problem · 1

PUMP IS RETURNED FOR MULTIPLE LOSS OF PRIME ALARMS. EVALUATION REVEALED A DAMAGED FORCE SENSOR ASSEMBLY. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 14 YR