FDA Adverse Event Malfunction Summary report: N

ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT

MDR report key: 21960163 · Received May 5, 2025

Report

Report Number
3005094123-2025-00215
Event Type
Malfunction
Date Received
May 5, 2025
Date of Event
April 23, 2025
Report Date
May 19, 2025
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
MMI
UDI-DI
00380740160630
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.

Additional Manufacturer Narrative · 0

THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW, IN HOUSE TESTING OF A RETAINED REAGENT KIT LOT 70021UD00 AND FIELD DATA REVIEW. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE WITHOUT INDICATION FOR ANY ADDITIONAL ISSUE. A SEARCH FOR SIMILAR COMPLAINTS IDENTIFIED AN INCREASE IN COMPLAINT ACTIVITY FOR LOT 70021UD00, HOWEVER, NO INCREASE IN COMPLAINT ACTIVITY WAS IDENTIFIED FOR LIST NUMBER 08P13. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES OR DEVIATIONS WITH THE COMPLAINT LOT. THE OVERALL PERFORMANCE OF THE ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENTS IN THE FIELD WAS REVIEWED USING DATA GATHERED FROM CUSTOMERS WORLDWIDE. THE PATIENT MEDIAN VALUES OBTAINED WITH THE COMPLAINT LOT 70021UD00 IS WITHIN ESTABLISHED LIMITS AND THUS COMPARABLE TO THE HISTORICAL LOT PERFORMANCE, WHICH CONFIRMS NO SYSTEMIC ISSUE FOR THE LOT. IN-HOUSE TESTING WAS COMPLETED USING PANELS WHICH MIMIC PATIENT SAMPLES USING AN IN-HOUSE RETAINED KIT. ALL SPECIFICATIONS WERE MET INDICATING THE LOT IS PERFORMING ACCEPTABLY. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER¿S ISSUE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT LOT 70021UD00 WAS IDENTIFIED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULT FOR ONE PATIENT. THE FOLLOWING INFORMATION WAS PROVIDED: (B)(6) 2025 AT 0949, SID 236255: INITIAL RESULT = 0.052 UG/L (B)(6) 2024 AT 1811, SID 230345: INITIAL RESULT WAS <0.01 UG/L (B)(6) 2024 AT 2104, SID 230372: INITIAL RESULT WAS <0.01 UG/L THE SAMPLES WERE REPEATED ON (B)(6) 2025 AND GENERATED RESULTS OF <0.01 UG/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I STAT HIGHLY SENSITIVE TROPONIN-I RESULT FOR ONE PATIENT. THE FOLLOWING INFORMATION WAS PROVIDED: (B)(6) 2025 AT 0949, SID (B)(6): INITIAL RESULT = 0.052 UG/L. (B)(6) 2024 AT 1811, SID (B)(6): INITIAL RESULT WAS <0.01 UG/L. (B)(6) 2024 AT 2104, SID (B)(6): INITIAL RESULT WAS <0.01 UG/L. THE SAMPLES WERE REPEATED ON 25APR2025 AND GENERATED RESULTS OF <0.01 UG/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
475490 ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ABBOTT IRELAND DIAGNOSTICS DIVISION 70021UD00 00380740160630

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ALNTY I PROCESSING MODU, 03R65-01, (B)(6) | ALNTY I PROCESSING MODU, 03R65-01, (B)(6)