Narratives

Additional Manufacturer Narrative · 0

E1 - INITIAL REPORTER ESTABLISHMENT NAME-(B)(6). E1 - ADDRESS- (B)(6). THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELDS: D8, G1, H2, H3, H6, H11. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: CONNECTION TEST NOT GOOD (NG) AND LEAK TEST NOT GOOD (NG) A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, THE MOST PROBABLE CAUSE OF THE MALFUNCTION DUE TO COMPONENT FAILURE. THREE ATTEMPTS WERE PERFORMED TO OBTAIN ADDITIONAL INFORMATION, BUT NO RESPONSE WAS RECEIVED FROM THE CUSTOMER. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SUBJECT DEVICE DISPLAYED AN ABNORMAL IMAGE. THIS ISSUE WAS IDENTIFIED DURING THE SETUP INSPECTION BEFORE THE PATIENT ENTERED THE ROOM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
422702 AUTOCLAVABLE CAMERA HEAD CAMERA HEAD FET SHIRAKAWA OLYMPUS CO., LTD. OTV-S7H-NA-12E

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown