FDA Adverse Event Malfunction Summary report: N

BD PYXIS¿ MEDSTATION¿ ES AUXILIARY

MDR report key: 21950240 · Received May 2, 2025

Report

Report Number
2016493-2025-77713
Event Type
Malfunction
Date Received
May 2, 2025
Date of Event
April 7, 2025
Report Date
May 2, 2025
Manufacturer
CAREFUSION 303, INC.
Product Code
BRY
UDI-DI
10885403512681
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE COMPLAINT HISTORY FOR (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINTS WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 13-APR-2022 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE CUBIES ON D AND C ROWS WERE DISPLAYED READ/WRITE ERRORS. A FIELD SERVICE ENGINEER LOGGED INTO HARDWARE TEST APPLICATION, OPENED AUXILIARY 1.1, POPPED OUT CUBIES AND CLEANED CONNECTIONS WITH ALCOHOL WIPES. REINSERTED CUBIES IN CORRECT LOCATION, POPPED OPEN LIDS. OPENED 1.2 AND POPPED OPEN ALL LIDS THERE ALSO. THEN, THE FIELD SERVICE ENGINEER REBOOTED THE STATION AND RECOVERED FAILED POCKETS TO RESOLVE THE ISSUE. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE FIELD SERVICE ENGINEER REPAIRED THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT WHEN USING THE BD PYXIS¿ MEDSTATION¿ ES AUXILIARY, THE CUBIES ON D AND C ROWS WERE DISPLAYED READ/WRITE ERRORS. THE CUSTOMER STATED THAT THIS MALFUNCTION CAUSED A DELAY IN DISPENSING THE MEDICATIONS TO THE PATIENT, BUT THE USER MANAGED TO RETRIEVE THE MEDICATION FROM ANOTHER DEVICE. HOWEVER, THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED DUE TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
209145 BD PYXIS¿ MEDSTATION¿ ES AUXILIARY AUTOMATED DISPENSING CABINET BRY CAREFUSION 303, INC. 324 10885403512681

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown