CERTAIN¿ TITANIUM LARGE HEXED SCREW
Report
- Report Number
- 0001038806-2025-00956
- Event Type
- Injury
- Date Received
- May 2, 2025
- Date of Event
- January 28, 2025
- Report Date
- July 31, 2025
- Manufacturer
- BIOMET 3I
- Product Code
- DZE
- PMA / PMN Number
- K972444
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER CMP-(B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT B5: DESCRIBE EVENT OR PROBLEM D4: ADDITIONAL DEVICE INFORMATION G3: DATE RECEIVED BY MANUFACTURER G6: TYPE OF REPORT H1: TYPE OF REPORTABLE EVENT H2: FOLLOW UP TYPE H3: DEVICE EVALUATED BY MANUFACTURER H4: DEVICE MANUFACTURER DATE H6: ADVERSE EVENT PROBLEM H10: ADDITIONAL NARRATIVE ZIMVIE RECEIVED ONE (1) ILRGHT, CERTAIN TITANIUM LARGE HEXED SCREW FOR EVALUATION. VISUAL EVALUATION WAS PERFORMED. FRACTURE SCREW IDENTIFIED INSIDE IMPLANT. DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1180734-CN. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1180734-CN FOR SIMILAR EVENTS AND ONE OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: FRACTURE: SCREW BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR. FRACTURE SCREW IDENTIFIED INSIDE IMPLANT. THE REPORTED EVENT WAS CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.
ZIMVIE COMPLAINT NUMBER (B)(4). A4: PATIENT WEIGHT: UNKNOWN / NOT PROVIDED, D4: ADDITIONAL DEVICE INFORMATION: UNKNOWN / NOT PROVIDED. D10: CONCOMITANT MEDICAL PRODUCT: ILRGHT, CERTAIN TITANIUM LARGE HEXED SCREW, LOT: 1180734-CN, E1: REPORTER NAME: UNKNOWN / NOT PROVIDED, G4: PREMARKET IDENTIFICATION PMA/510(K) #: K072642, H4: DEVICE MANUFACTURER DATE: UNKNOWN / NOT PROVIDED.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
DOCTOR REPORTED THAT PATIENT CAME TO CLINIC REFERRING MOBILITY OF CROWNS, DOCTOR NOTICED SCREW FRACTURE AT TOOTH SITES 36 AND 37. WHEN SCREWS COULD NOT BE REMOVED, IMPLANTS WERE REMOVED. NEW IMPLANTS WILL BE PLACED IN 9 MONTHS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 189473 | CERTAIN¿ TITANIUM LARGE HEXED SCREW | DENTAL SCREW | DZE | BIOMET 3I | 1180734-CN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Male | Required Intervention | DENTAL SCREW SEE H10 NARRATIVE. |