FDA Adverse Event Malfunction Summary report: N

PERMAFLEX PLUS ADHESIVE SLIDES, WHITE

MDR report key: 21947197 · Received May 1, 2025

Report

Report Number
1419341-2025-00027
Event Type
Malfunction
Date Received
May 1, 2025
Date of Event
April 3, 2025
Report Date
May 30, 2025
Manufacturer
LEICA BIOSYSTEMS RICHMOND INC.
Product Code
KEW
UDI-DI
20849832023146
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER TESTING OF RETAIN SAMPLES WAS UNABLE TO REPLICATE THE "TISSUE IS FALLLNG [SIC] OFF THE SLIDES DURING IHC" REPORTED BY THE CUSTOMER. THE ROOT CAUSE FOR THE "TISSUE IS FALLLNG [SIC] OFF THE SLIDES DURING IHC" REPORTED BY THE CUSTOMER COULD NOT BE UNEQUIVOCALLY DETERMINED FROM THE INFORMATION AVAILABLE. NO ADVERSE CONSEQUENCE(S) TO A PATIENT(S) HAS BEEN REPORTED TO THE MANUFACTURER TO DATE. IF ADDITIONAL INFORMATION IS RECEIVED AN ADDITIONAL FOLLOW UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

DEVICE HISTORY REVIEW WAS COMPLETED. THERE WERE NO DEVIATIONS OR NON-CONFORMANCES ASSOCIATED WITH THIS LOT. TRENDING ANALYSIS REPORTED NO (0) OTHER RELATED OCCURRENCE OF THIS ISSUE FOR THIS PRODUCT LOT. QUALITY CONTROL TESTING WAS COMPLETE AND PASSED. IF ADDITIONAL INFORMATION IS RECEIVED AN ADDITIONAL FOLLOW UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 0

ON 03 APRIL 2025, LEICA BIOSYSTEMS RECEIVED THE FOLLOWING INFORMATION: "TISSUE IS FALLING OFF THE SLIDES; 2ND SECTION FALLS OFF, OVEN IS USED 30 MIN, AFFECTS ONLY IHC, NOT ALL SECTIONS FALL OFF. NO IRRETRIEVABLE TISSUE LOSS WAS REPORTED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1023143 PERMAFLEX PLUS ADHESIVE SLIDES, WHITE SLIDES, MICROSCOPE KEW LEICA BIOSYSTEMS RICHMOND INC. PERMAFLEX PLUS ADHESIVE SLIDES, WHITE 022525-1 20849832023146

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown