FDA Adverse Event Injury Summary report: N

EVOLIS TIBIAL BASE PLATE TI PLASMA SPRAYED

MDR report key: 2194682 · Received July 29, 2011

Report

Report Number
3005180920-2011-00043
Event Type
Injury
Date Received
July 29, 2011
Date of Event
July 19, 2011
Report Date
July 29, 2011
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
PMA / PMN Number
K081023
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DOCUMENT REVIEW: EVOLIS TI PLASMA SPRAYED BASEPLATE - (B)(4) / LOT. 090155. THE ANALYSIS OF THE BATCH RECORDS WAS DONE AND NO ANOMALIES WERE FOUND CONCERNING THE PROBLEM OCCURRED. EVOLIS TI PLASMA SPRAYED BASEPLATE IN THE USA IS CLEARED UNDER K081023 (EVOLIS TOTAL KNEE SYSTEM) AS CEMENTED PROSTHESIS, BUT THE SURGEON IMPLANTED IT W/O CEMENT. IN THE IFU AND SURGICAL TECHNIQUE FOR US OF EVOLIS, IS CLEARLY INDICATED THAT THE PRODUCT IS TO BE CEMENTED. W/O USING CEMENT, IT IS KNOWN THAT THE RISK OF LOOSENING IS HIGHER THAN CEMENTED SYSTEMS. THE EVENT IS NOT DEVICE RELATED, BUT DUE TO A USE ERROR.

Description of Event or Problem · 1

TIBIAL BASEPLATE WAS NOT FIXED IN PLACE AND CAME LOOSE. A REVISION SURGERY WAS NECESSARY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EVOLIS TIBIAL BASE PLATE TI PLASMA SPRAYED TIBIAL BASE PLATE SIZE 3 JWH MEDACTA INTERNATIONAL SA 090155

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention