FDA Adverse Event
Malfunction
Summary report: N
PANORAMA
MDR report key: 2193620
·
Received July 14, 2011
Report
- Report Number
- 2221819-2011-00006
- Event Type
- Malfunction
- Date Received
- July 14, 2011
- Date of Event
- May 13, 2011
- Report Date
- July 30, 2011
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- MHX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY SERVICE REPRESENTATIVE REPLACED THE POWER SUPPLY IN THE WIRELESS TRANSCEIVER. THE SYSTEM WAS TESTED TO FACTORY SPECIFICATIONS. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT WHILE THE PANORAMA CENTRAL STATION WAS IN USE MONITORING PATIENTS ALONG WITH SPECTRUM MONITORS AND AMBULATORY TELEPACKS, THE WIRELESS TRANSCEIVER STOPPED FUNCTIONING, CAUSING THE CENTRAL STATION TO LOSE CONNECTIVITY TO THE PATIENT DEVICES. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PANORAMA | CENTRAL STATION MONITORING NETWORK | MHX | MINDRAY DS USA, INC. | PANORAMA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |