FDA Adverse Event Malfunction Summary report: N

MPS DELIVERY SET W/ARREST AGENT, ADDITIVE CASS

MDR report key: 2193599 · Received July 14, 2011

Report

Report Number
1649914-2011-00008
Event Type
Malfunction
Date Received
July 14, 2011
Date of Event
June 15, 2011
Report Date
June 16, 2011
Manufacturer
QUEST MEDICAL
Product Code
DTR
PMA / PMN Number
953838
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THE PERFUSIONIST OBSERVED FLUID HAD LEAKED FROM THE ARREST CASSETTE OF THE DELIVERY SET AND AS A RESULT HE WAS UNABLE TO SUCCESSFULLY ARREST THE PT'S HEART DURING THE PROCEDURE. THE PERFUSIONIST STATED HE THOUGHT THE FLUID APPEARED TO HAVE LEAKED DOWN INTO THE MAIN CONSOLE UNIT. AS A PRECAUTION, THE CONSOLE WAS NOT USED TO COMPLETE THE PROCEDURE. A MANUAL METHOD USING A ROLLER PUMP WAS USED INSTEAD AND THE PROCEDURE WAS COMPLETED WITH NO PT COMPLICATIONS REPORTED. THE DELIVERY SET WAS NOT RETURNED TO THE MFR FOR ANALYSIS. NO FURTHER INFO IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MPS DELIVERY SET W/ARREST AGENT, ADDITIVE CASS CPBP HEAT EXCHANGER DTR QUEST MEDICAL 5001102

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention