FDA Adverse Event Malfunction Summary report: N

GEMSTAR PM THERAPY S

MDR report key: 2193580 · Received July 14, 2011

Report

Report Number
9615050-2011-00533
Event Type
Malfunction
Date Received
July 14, 2011
Date of Event
June 20, 2011
Report Date
June 21, 2011
Manufacturer
HOSPIRA INC.
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT THE PUMP DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FROM THE LABOR AND DELIVERY DEPARTMENT WITH A REPORT OF "10/000/000, NO ALARM AND INOPERABLE." NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAY OF CRITICAL THERAPIES WHEN THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GEMSTAR PM THERAPY S 80FRN FRN HOSPIRA INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK