ACCURUS 800CS
Report
- Report Number
- 2028159-2011-00805
- Event Type
- Malfunction
- Date Received
- July 13, 2011
- Report Date
- March 16, 2011
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CTR
- Product Code
- HQC
- PMA / PMN Number
- K911808
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
COMPLAINT EVAL PERFORMED ON THE RETURNED SAMPLE (23GA ACCURUS) RESULTED IN THE FOLLOWING FINDINGS: A VISUAL INSPECTION DEEMED THE SAMPLE WAS CONFORMING. THE SAMPLE WOULD ACTUATE PROPERLY, BUT WOULD NOT CUT PROPERLY (STRANDS). THE PROBE ASPIRATED ACCORDINGLY. THE DEVICE HISTORY RECORDS FOR THE COMPONENT LOTS WERE REVIEWED. THE ASSOCIATED LOTS (9) WERE RELEASED BASED ON COMPANY'S ACCEPTANCE CRITERIA. THE PROBE WAS DISASSEMBLED AND THE COMPONENTS INSPECTED. THE INNER CUTTER EXHIBITED SIGNIFICANT WEAR ON THE CUTTING EDGE (RESHAPED) INDICATING THE PROBE MAY HAVE BEEN USED EXTENSIVELY. THE SIGNIFICANTLY WORN/DAMAGED INNER CUTTER EDGE HAS AFFECTED THE CUTTING ABILITY OF THE PROBE. IT MUST ALSO BE NOTED THAT MACHINE VIT VALVES NEEDED TO BE REPAIRED WHEN IT WAS EVALUATED BY FIELD SERVICE PERSONNEL. THE PRODUCT EVAL DETERMINED A DAMAGED CUTTING EDGE. SIGNIFICANT WEAR AND NICK ON THE CUTTING EDGE INDICATES THAT THE PROBE MAY HAVE BEEN INITIALLY WORKING PROPERLY AND ONLY DAMAGED SOMETIME DURING SURGERY. SIGNIFICANT USE OF THE PROBE WAS EVIDENT. FURTHERMORE, THE MACHINE'S VIT VALVES NEEDED TO BE REPAIRED WHEN IT WAS EVALUATED BY FIELD SERVICE PERSONNEL. (B)(4).
A CUSTOMER REPORTED THE VITRECTOMY PROBES WERE NOT CUTTING AND THE ASPIRATION WOULD NOT FUNCTION. THE CUSTOMER INDICATED ALL THE PROBES ARE TESTED TWICE PRIOR TO USE; ONCE BY THE SCRUB TECH, AND ONCE BY THE SURGEON BEFORE ENTERING THE PT'S EYE. THERE HAVE BEEN NO PTS HARMED BY THESE PROBES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCURUS 800CS | PHACOFRAGMENTATION SYSTEM | HQC | ALCON - IRVINE TECHNOLOGY CTR | ACCURUS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | TOTAL PLUS PACK, 23 GAUGE |