FDA Adverse Event Malfunction Summary report: N

SEE H-10

MDR report key: 2193412 · Received August 5, 2011

Report

Report Number
2023826-2011-00696
Event Type
Malfunction
Date Received
August 5, 2011
Date of Event
July 12, 2011
Report Date
July 12, 2011
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P990013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND THE LENS TORN INTO TWO PIECES, WITH PIECES FROM BOTH HAPTICS TORN OFF AND MISSING. THE LENS WAS RETURNED IN LIQUID. THE TIP OF THE RETURNED CARTRIDGE WAS DAMAGED. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. CONCLUSIONS - (NO CONCLUSION CAN BE DRAWN):BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND THE EVALUATION OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED.(B)(4).

Additional Manufacturer Narrative · 1

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCULAR LENS. (B)(4) - LENS (IOL), TORN, SPLIT, CRACKED. EVALUATION METHOD: LENS WORK ORDER SEARCH. RESULTS: A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. CONCLUSIONS: (NO CONCLUSION CAN BE DRAWN): BASED ON THE COMPLAINT HISTORY AND WORK ORDER SEARCH, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4). LENS NOT RETURNED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION RECEIVED - THE LENS WAS TORN BY THE INJECTION SYSTEM. THE SURGEON USED A SUTURE TO SEAL THE INCISION.

Description of Event or Problem · 1

THE REPORTER STATED THE SURGEON INSERTED A CC4204A COLLAMER ASPHERIC SINGLE PIECE LENS AND THE LENS TORE. THE LENS WAS REMOVED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SEE H-10 INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY CC4204A NA

Patients

Seq Age Sex Outcome Treatment
1 85 YR INJECTION SYSTEM LOT NUMBER 101405| INJECTION SYSTEM MODEL NANOPOINT - LOT NUMBER UNK