FDA Adverse Event Injury Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 2193363 · Received July 29, 2011

Report

Report Number
2032227-2011-01862
Event Type
Injury
Date Received
July 29, 2011
Date of Event
July 14, 2011
Report Date
July 14, 2011
Manufacturer
MEDTRONIC MINIMED
Product Code
LZG
PMA / PMN Number
K040676
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS, WITH A BLOOD GLUCOSE READING OF 540 MG/DL. THE CUSTOMER WAS VOMITING PRIOR TO THE EVENT. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP PASSED THE PRIME TEST, BUT FAILED THE HIGH PRESSURE TEST. THE CUSTOMER STATED THAT HE HAS INSERTED THE INFUSION SETS IN HIS ABDOMEN FOR FIVE YEARS, AND HAS NOT BEEN CHECKED FOR SCAR TISSUE. ADVISED THE CUSTOMER TO GET CHECKED FOR SCAR TISSUE. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG MEDTRONIC MINIMED MMT-715LNAS

Patients

Seq Age Sex Outcome Treatment
1 29 YR Hospitalization (B)(4) QUICK-SET PARADIGM INSULIN INFUSION SET| SILHOUETTE