PARADIGM INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2011-01862
- Event Type
- Injury
- Date Received
- July 29, 2011
- Date of Event
- July 14, 2011
- Report Date
- July 14, 2011
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- LZG
- PMA / PMN Number
- K040676
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS, WITH A BLOOD GLUCOSE READING OF 540 MG/DL. THE CUSTOMER WAS VOMITING PRIOR TO THE EVENT. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP PASSED THE PRIME TEST, BUT FAILED THE HIGH PRESSURE TEST. THE CUSTOMER STATED THAT HE HAS INSERTED THE INFUSION SETS IN HIS ABDOMEN FOR FIVE YEARS, AND HAS NOT BEEN CHECKED FOR SCAR TISSUE. ADVISED THE CUSTOMER TO GET CHECKED FOR SCAR TISSUE. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | MEDTRONIC MINIMED | MMT-715LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR | Hospitalization | (B)(4) QUICK-SET PARADIGM INSULIN INFUSION SET| SILHOUETTE |