FDA Adverse Event Injury Summary report: N

CONTOUR

MDR report key: 2193348 · Received July 28, 2011

Report

Report Number
1826988-2011-00441
Event Type
Injury
Date Received
July 28, 2011
Date of Event
July 9, 2011
Report Date
July 9, 2011
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER DID NOT PROVIDE ANY PRODUCT INFO. IT'S NOT POSSIBLE TO DETERMINE THE MANUFACTURE DATE OR 510(K) NUMBER WITHOUT THE METER INFO. THE CUSTOMER'S USB CABLE WAS RETURNED AND THE CASE CONTAINING THE PRINTED CIRCUIT BOARD ASSEMBLY (PCBA) WAS VERIFIED TO HAVE BEEN MELTED. THE CABLE WAS TESTED WITH 3 DIFFERENT METERS AND CONNECTED TO A COMPUTER OVER A 30 MINUTE TIME PERIOD. THE CABLE WORKED CORRECTLY AND DOWNLOADED ALL 3 METERS. THE CABLE DID NOT GET HOT DURING THE 30 MINUTES. THE CASE AROUND THE PCBA WAS CUT OPEN. THERE WERE SIGNS OF OVERHEATING. SOME COMPONENTS MELTED OR CAME OFF THE BOARD, WHICH CAUSED THE PLASTIC HOUSING TO MELT. THE CABLE IS TO BE RETURNED TO THE SUPPLIER FOR FURTHER EVAL.

Description of Event or Problem · 1

A CUSTOMER FROM (B)(6) CONTACTED CUSTOMER SVC. HE INSTALLED THE GLUCOFACTS SOFTWARE ON HIS COMPUTER AND THEN CONNECTED HIS CONTOUR METER USING A BAYER USB CABLE. THE SOFTWARE DID NOT APPEAR TO DETECT THE METER, SO THE CUSTOMER CHECKED THE CABLE CONNECTION. WHEN HE TOUCHED THE CABLE, IT WAS VERY HOT AND THE JUNCTION BOX ON THE CABLE HAD MELTED SLIGHTLY. THE CUSTOMER STATED HIS CARPET WAS SLIGHTLY BURNT AND HE ALSO REC'D A BLISTERING BURN ON HIS HAND. THE CABLE WAS RETURNED FOR EVAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR NBW BAYER HEALTHCARE, LLC NOT PROVIDED NA

Patients

Seq Age Sex Outcome Treatment
1 UNK