TRIAGE PROFILER SOB 25 TEST KIT
Report
- Report Number
- 2027969-2011-01673
- Event Type
- Injury
- Date Received
- July 28, 2011
- Date of Event
- June 28, 2011
- Report Date
- July 28, 2011
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- NBC
- PMA / PMN Number
- K042723
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT SUPPORT TESTED CAL H (POS CTRL) AND CAL Z (NEG CTRL) ON SOB LOT W48946. OBSERVATIONS WERE FOR TNI RECOVERY. NO HIGH BIAS OR FALSE POSITIVE RESULTS WERE OBSERVED WITH ANY SAMPLE. CUSTOMER DID NOT RETURN ANY SAMPLE FOR FURTHER TESTING. UNABLE TO RULE OUT SAMPLE SPECIFIC INTERFERENCE THAT MAY HAVE AFFECTED ANALYTE RECOVERY. ALL RESULTS FOR TNI GIVEN BY CUSTOMER ARE BELOW THE CLINICAL CUTOFF OF 0.40NG/ML AS STATED IN THE PACKAGE INSERT. NO PRODUCT DEFICIENCY WAS ESTABLISHED. AS OF (B)(6) 2011 THERE ARE 3 COMPLAINTS FOR SOB LOT W48946. CORRECTIVE ACTION NOT REQUIRED AT THIS TIME.
CALLER ALLEGED FALSE POSITIVE TROPONIN (TNI) RESULTS USING THE TRIAGE PROFILER SOB. PT F: PT WENT FOR AN URGENT CARDIAC CATHETERIZATION WITH NORMAL CORONARIES. CT CHEST SHOWED NO PULMONARY EMBOLISM. CARDIAC ECHOCARDIOGRAM WAS WITHIN NORMAL LIMITS. DISCHARGED (B)(6) 2011 WITH DIABETES MELLITUS, HYPOTHYROIDISM, HYPERTENSION, DEPRESSION AND ATYPICAL CHEST PAIN. THE FOLLOWING CUT OFF'S WERE USED: TNI: 0.01-0.05-BORDERLINE. NTPRO BNP: >450 (50Y) >900 (50-76Y) >1800 (>76Y). CK: 20-200, CKMB: 02.-2.4, DDIM: 0-400.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRIAGE PROFILER SOB 25 TEST KIT | CARDIAC MARKER TEST | NBC | ALERE SAN DIEGO, INC. | 97300 | W48946B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |