FDA Adverse Event Injury Summary report: N

TORNIER PYROCARBON HUM HEAD DIA 52MMX19MMX1.5MM ECC LO

MDR report key: 21933337 · Received April 30, 2025

Report

Report Number
3000931034-2025-00233
Event Type
Injury
Date Received
April 30, 2025
Date of Event
April 4, 2025
Report Date
July 22, 2025
Manufacturer
TORNIER S.A.S.
Product Code
KWS
UDI-DI
037003886943355
PMA / PMN Number
DEN220012
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ONCE THE INVESTIGATION IS COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT

Additional Manufacturer Narrative · 0

THE REPORTED EVENT COULD NOT BE CONFIRMED SINCE THE DEVICE WAS NOT RETURNED FOR EVALUATION AND NO OTHER EVIDENCES WERE PROVIDED. MORE DETAILED INFORMATION ABOUT THE COMPLAINT EVENT AS WELL AS THE AFFECTED DEVICE MUST BE AVAILABLE IN ORDER TO DETERMINE THE ROOT CAUSE OF THE COMPLAINT EVENT. A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. NO INDICATIONS OF MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE FOUND DURING THE INVESTIGATION. IF THE DEVICE IS RETURNED OR IF ANY ADDITIONAL INFORMATION IS PROVIDED, THE INVESTIGATION WILL BE REASSESSED.

Description of Event or Problem · 0

PT UNDERWENT LEFT SHOULDER HEMI-ARTHROPLASTY, ON (B)(6) 2018. A PYRO-CARBON HEAD WAS USED INITIALLY. RECENTLY PT HAS BEEN COMPLAINING OF INCREASING DISCOMFORT IN THIS JOINT AND PRESENTED FOR REVIEW. SURGEON PERFORMED A SHOULDER ARTHROSCOPY ON THIS JOINT WITH MILD GLENOID ARTHRITIS NOTED. PT INSISTED ON HAVING A SHOULDER REVISION (GLENOID RESURFACING), EVEN THOUGH SURGEON RECOMMENDED MORE CONSERVATIVE MANAGEMENT (ORAL ANALGESICS). DURING SURGERY, MILD GLENOID ARTHRITIC CHANGES WERE NOTED, AND SURGEON DECIDED TO GO AHEAD, AS PER PATIENT'S WISHES, AND RESURFACE THE GLENOID - A POLY GLENOID PROSTHESIS WAS IMPLANTED. PYROCARBON HEAD REPLACED WITH A METAL HEAD, AS THE PYC HEAD IS NOT INDICATED FOR USE WITH A POLY GLENOID COMPONENT. JOINT STABLE AFTER REDUCTION. WOUND CLOSED.

Description of Event or Problem · 0

PT UNDERWENT LEFT SHOULDER HEMI-ARTHROPLASTY, ON (B)(6) 2018. A PYRO-CARBON HEAD WAS USED INITIALLY. RECENTLY PT HAS BEEN COMPLAINING OF INCREASING DISCOMFORT IN THIS JOINT AND PRESENTED FOR REVIEW. SURGEON PERFORMED A SHOULDER ARTHROSCOPY ON THIS JOINT WITH MILD GLENOID ARTHRITIS NOTED. PT INSISTED ON HAVING A SHOULDER REVISION (GLENOID RESURFACING), EVEN THOUGH SURGEON RECOMMENDED MORE CONSERVATIVE MANAGEMENT (ORAL ANALGESICS). DURING SURGERY, MILD GLENOID ARTHRITIC CHANGES WERE NOTED, AND SURGEON DECIDED TO GO AHEAD, AS PER PATIENT'S WISHES, AND RESURFACE THE GLENOID - A POLY GLENOID PROSTHESIS WAS IMPLANTED. PYROCARBON HEAD REPLACED WITH A METAL HEAD, AS THE PYC HEAD IS NOT INDICATED FOR USE WITH A POLY GLENOID COMPONENT. JOINT STABLE AFTER REDUCTION. WOUND CLOSED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1413123 TORNIER PYROCARBON HUM HEAD DIA 52MMX19MMX1.5MM ECC LO PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED KWS TORNIER S.A.S. 6418AS 037003886943355

Patients

Seq Age Sex Outcome Treatment
1 56 YR Male Required Intervention