FDA Adverse Event Injury Summary report: N

SMR SHOULDER

MDR report key: 21932973 · Received April 30, 2025

Report

Report Number
3008021110-2025-00040
Event Type
Injury
Date Received
April 30, 2025
Date of Event
March 24, 2025
Report Date
April 30, 2025
Manufacturer
LIMACORPORATE S.P.A.
Product Code
KWT
UDI-DI
08033390001922
PMA / PMN Number
K100858
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: CHECKING THE MANUFACTURING CHARTS OF THE COMPONENTS INVOLVED IN THE EVENT, NO PRE-EXISTING ANOMALY WAS FOUND ON THE ITEMS BELONGING TO THE SAME PART CODES AND LOT NUMBERS AS THE COMPONENTS REMOVED DURING THE REVISION SURGERY HEREBY REPORTED. WE WILL SUBMIT A FINAL MDR AS SOON AS THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 0

REVISION SURGERY PERFORMED ON (B)(6) 2025, DUE TO DISASSEMBLY OF COMPONENTS. THE MORSE TAPER CONNECTION DISENGAGED. THE FOLLOWING COMPONENTS WERE REMOVED: SMR FINNED STEM SHORT D.21 (PART CODE: 1304.15.021, LOT NUMBER: 2016623, STERILIZATION 2000409). SMR HUMERAL HEAD DIA.54 MM (PART CODE: 1322.09.540, LOT NUMBER: 1506372, STERILIZATION 2000414). SMR ECC.ADAPTOR TAPER STANDARD (PART CODE: 1330.15.272, LOT NUMBER: 2331971, STERILIZATION 2400033). SMR FINNED HUMERAL BODY (PART CODE: 1350.15.110, LOT NUMBER: 2101472, STERILIZATION 2100098). THE ABOVE-MENTIONED COMPONENTS WERE REPLACED BY THE FOLLOWING NEW ONES: SMR FINNED STEM DIA 21MM L.80 (PART CODE: 1304.15.210, LOT NUMBER: 2320690, STERILIZATION 2400025). SMR FINNED HUMERAL BODY (PART CODE: 1350.15.110, LOT NUMBER: 2405095, STERILIZATION 2400062). SMR ECC. ADAPTOR TAPER 2MM (PART CODE: 1330.15.272, LOT NUMBER: 2414583, STERILIZATION 2400152). SMR HUMERAL HEAD DIA.54 MM (PART CODE: 1322.09.540, LOT NUMBER: 1503494, STERILIZATION 2000263). PREVIOUS SURGERY TOOK PLACE ON (B)(6) 2025. THE PATIENT IS A MALE, DATE OF BIRTH ON (B)(6) 1972. THE EVENT HAPPENED IN THE UNITED STATES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1581714 SMR SHOULDER ECCENTRICAL ADAPTOR TAPERS +2MM, STANDARD KWT LIMACORPORATE S.P.A. 1330.15.272 2331971 08033390001922

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention