FDA Adverse Event Malfunction Summary report: N

HEARTMATE II SYSTEM CONTROLLER

MDR report key: 2193284 · Received July 13, 2011

Report

Report Number
2916596-2011-00293
Event Type
Malfunction
Date Received
July 13, 2011
Date of Event
June 7, 2011
Report Date
June 14, 2011
Manufacturer
THORATEC CORP.
Product Code
DSQ
Removal / Correction Number
29165969/2/10001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MFR FOR EVAL. THE SYSTEM CONTROLLER WAS CONNECTED TO THE EQUIPMENT IN THE MFR'S LAB WITH NO ALARMS ACTIVE. THE BLACK AND WHITE POWER LEADS WERE MANIPULATED AND WHILE MOVING THE WHITE POWER LEAD AT THE CONNECTOR END, THE SYSTEM REPORTED A "LOW BATTERY" ALARM FOLLOWED BY A "POWER CABLE DISCONNECTED" ALARM. THE SYSTEM CONTROLLER WAS REMOVED FROM THE TEST EQUIPMENT. THE OUTER INSULATION AND THE STRAIN RELIEF OF THE WHITE POWER LEAD AT THE CONNECTOR END WAS REMOVED REVEALING A BROKEN INNER CONDUCTOR. THIS SITUATION IS BEING ADDRESSED THROUGH THE MFR'S CORRECTIVE/PREVENTIVE ACTION SYSTEM AND AN URGENT MEDICAL DEVICE CORRECTION NOTICE (29165969/2/10001-C) WAS SENT TO CUSTOMERS. NO FURTHER INFO IS AVAILABLE. THE MFR IS CLOSING ITS FILE ON THIS EVENT.

Description of Event or Problem · 1

THE PT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD). THE TRANSPLANT COORDINATOR REPORTED THAT THE VAD COORDINATOR RECEIVED A PHONE CALL FROM THE PT NOTING THAT HE WAS HAVING ONGOING INTERMITTENT ALARMS FROM HIS SYSTEM CONTROLLER. THE ALARMS WERE ASSOCIATED WITH THE FLASHING OF THE BATTERY LIGHTS. HE WAS ATTEMPTING TO TEST THE BATTERIES ON THE SYSTEM CONTROLLER, AND THEY TESTED FULL EVEN THOUGH THEY WERE COMPLETELY DEPLETED. THE PT STATED THAT OVERALL HE WAS FEELING WELL. IT WAS REPORTED THAT THE LVAD PARAMETERS WERE NORMAL WITH SPEED AT 10,200 RPMS. THE PT HAS BEEN COMPLETING NORMAL ACTIVITY AND HAS NOTED NO CHANGE IN HIS LVAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTMATE II SYSTEM CONTROLLER LVAD SYSTEM CONTROLLER DSQ THORATEC CORP. 103696 90899

Patients

Seq Age Sex Outcome Treatment
1 59 YR Other