FDA Adverse Event Injury Summary report: N

CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS

MDR report key: 2193267 · Received August 5, 2011

Report

Report Number
2015691-2011-15992
Event Type
Injury
Date Received
August 5, 2011
Date of Event
June 9, 2011
Report Date
July 8, 2011
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DYE
PMA / PMN Number
P860057/S018
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) DEVICE NOT RETURNED. ADDITIONAL MANUFACTURER NARRATIVE: UNFORTUNATELY, THE EDWARDS DEVICE WAS NOT RETURNED FOR ANALYSIS. OTHER ADDITIONAL INFORMATION (E.G. OPERATIVE REPORT, DISCHARGE SUMMARY, ETC.) WAS ALSO REQUESTED, HOWEVER, IT HAS YET TO BE PROVIDED. CALCIFICATION IS A WELL RECOGNIZED FAILURE MODE OF BIOPROSTHETIC VALVES. THE MECHANISMS FOR BIOPROSTHETIC HEART VALVE TISSUE CALCIFICATION ARE NOT FULLY UNDERSTOOD. MANY FACTORS CAN CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION INCLUDING PATIENT RELATED (E.G. PATIENT AGE, DISEASE STATE, IMMUNE STATUS, AND OTHER CO-MORBIDITIES), PHARMACOLOGICAL, AND INTRINSIC PROPERTIES OF THE VALVE ITSELF. IT IS WIDELY UNDERSTOOD THAT PATIENTS WITH CHRONIC RENAL DISEASE AND PRIOR HISTORY OF CALCIFIC STENOSIS OF THE NATIVE VALVE MAY BE PREDISPOSED TO BIOPROSTHETIC CALCIFICATION.

Additional Manufacturer Narrative · 1

THROUGH FOLLOW-UP WITH THE HEALTH-CARE PROVIDER, THE ADDITIONAL INFORMATION WAS RECEIVED. ACCORDING TO THE OPERATIVE REPORT, THE PATIENT DEVELOPED CRITICAL AORTIC STENOSIS IN HIS PROSTHETIC VALVE. THE VALVE WAS FOUND TO BE SEVERELY CALCIFIED, INCLUDING THE LEAFLETS WITH VERY LIMITED MOBILITY. THE DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED AND THIS DEVICE PASSED ALL MANUFACTURING AND STERILIZATION INSPECTIONS WITH NO NONCONFORMANCE.

Description of Event or Problem · 1

EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL "CUSTOMER COMPLAINT". THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE. IN THIS CASE, IT WAS REPORTED THAT THE DEVICE WAS EXPLANTED AFTER AN IMPLANT DURATION OF APPROXIMATELY 5 YEARS. THROUGH FOLLOW-UP WITH THE HEALTH-CARE PROVIDER, IT WAS LEARNED THAT THE DEVICE WAS EXPLANTED DUE TO AORTIC STENOSIS AND CALCIFICATION. UNFORTUNATELY, NO OTHER DETAILS WERE PROVIDED. ACCORDING TO OUR RECORDS, THE DEVICE WAS REPLACED WITH ANOTHER EDWARDS BIOPROSTHETIC VALVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS REPLACEMENT HEART VALVE DYE EDWARDS LIFESCIENCES 3000 R-06G0062

Patients

Seq Age Sex Outcome Treatment
1 64 YR Hospitalization| R