FLEXCATH STEERABLE SHEATH
Report
- Report Number
- 3002648230-2011-00118
- Event Type
- Malfunction
- Date Received
- August 5, 2011
- Date of Event
- July 11, 2011
- Report Date
- July 11, 2011
- Manufacturer
- MEDTRONIC CRYOCATH LP
- Product Code
- DRA
- PMA / PMN Number
- K081049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE WAS RETURNED AND WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED ISSUE WAS CONFIRMED THROUGH TESTING. DISSECTION SHOWED A LEAKING HEMOSTATIC VALVE. A FIELD ACTION WAS INITIATED IN JULY 2011 TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THIS POTENTIAL LEAK IN HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE DEVICE HAS NOT YET BEEN RETURNED FOR EVALUATION. RESULTS OF EVALUATION OF RETURNED DEVICE WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.
THE FLEXCATH STEERABLE SHEATH WAS PLACED IN THE PATIENT AND THE ARCTIC FRONT CATHETER WAS INSERTED INTO THE SHEATH. ON SLOW ASPIRATION, A CONTINUOUS DRAW OF AIR WAS ASPIRATED. A TOTAL OF APPROXIMATELY 40 CC OF AIR WAS ASPIRATED. THE SHEATH WAS REPLACED AND NO ADVERSE EVENT OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLEXCATH STEERABLE SHEATH | CATHETER, STEERABLE | DRA | MEDTRONIC CRYOCATH LP | 3FC12 | 35329 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |