FDA Adverse Event Malfunction Summary report: N

FLEXCATH STEERABLE SHEATH

MDR report key: 2193204 · Received August 5, 2011

Report

Report Number
3002648230-2011-00118
Event Type
Malfunction
Date Received
August 5, 2011
Date of Event
July 11, 2011
Report Date
July 11, 2011
Manufacturer
MEDTRONIC CRYOCATH LP
Product Code
DRA
PMA / PMN Number
K081049
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED AND WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED ISSUE WAS CONFIRMED THROUGH TESTING. DISSECTION SHOWED A LEAKING HEMOSTATIC VALVE. A FIELD ACTION WAS INITIATED IN JULY 2011 TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THIS POTENTIAL LEAK IN HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT YET BEEN RETURNED FOR EVALUATION. RESULTS OF EVALUATION OF RETURNED DEVICE WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

THE FLEXCATH STEERABLE SHEATH WAS PLACED IN THE PATIENT AND THE ARCTIC FRONT CATHETER WAS INSERTED INTO THE SHEATH. ON SLOW ASPIRATION, A CONTINUOUS DRAW OF AIR WAS ASPIRATED. A TOTAL OF APPROXIMATELY 40 CC OF AIR WAS ASPIRATED. THE SHEATH WAS REPLACED AND NO ADVERSE EVENT OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXCATH STEERABLE SHEATH CATHETER, STEERABLE DRA MEDTRONIC CRYOCATH LP 3FC12 35329

Patients

Seq Age Sex Outcome Treatment
1