FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2193190 · Received July 22, 2011

Report

Report Number
1720753-2011-20334
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
July 8, 2011
Report Date
July 22, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE INTERCONNECT CABLE, THE BATTERY CHARGER PRINTED CIRCUIT BOARD, AND THE POWER SIGNAL INTERFACE PRINTED CIRCUIT BOARD WERE REPLACED. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SVC.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED AN ERROR MESSAGE AND WOULD NOT BOOT UP. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1