FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 2193143 · Received July 22, 2011

Report

Report Number
1720753-2011-20298
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
July 11, 2011
Report Date
July 22, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE SRAM CARD WAS REPLACED. THE BOOST DOSE RATE WAS SET. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SVC.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYS DISPLAYED A CHECK SUM ERROR MESSAGE AND THE BOOT UP PROCESS WAS TERMINATED. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1