FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 2193125 · Received July 22, 2011

Report

Report Number
1644487-2011-01671
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
March 1, 2011
Report Date
July 18, 2011
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PHYSICIAN WAS HAVING PROBLEMS WITH INTERMITTENT COMMUNICATION WITH HIS PROGRAMMING SYSTEM AND HE ATTRIBUTED THE PROBLEM TO THE PROGRAMMING WAND. HOWEVER, THE PROGRAMMING SYSTEM WAS RETURNED TO MFR FOR ANALYSIS AND DURING ANALYSIS, A PROBLEM WAS FOUND WITH THE HANDHELD COMPUTER. DURING THE ANALYSIS IT WAS IDENTIFIED THAT THE HANDHELD WAS UNABLE TO ESTABLISH COMMUNICATION WITH A KNOWN GOOD GENERATOR. THE CAUSE OF THE ANOMALY IS ASSOCIATED WITH A DAMAGED GROUND CONNECTION IN THE SERIAL CABLE. ONCE THE GROUND CONNECTION WAS RE-ESTABLISHED USING A JUMPER WIRE, NO FURTHER ANOMALIES WERE IDENTIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. 250 715914

Patients

Seq Age Sex Outcome Treatment
1