FDA Adverse Event
Malfunction
Summary report: N
2800
MDR report key: 2193113
·
Received July 22, 2011
Report
- Report Number
- 1720753-2011-20284
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- July 11, 2011
- Report Date
- July 22, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE GE SERVICE REP CONDUCTED AN ONSITE INVESTIGATION. THE ION CHAMBER WAS REPLACED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A GENERATOR ERROR MESSAGE. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 2800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 2800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |