FDA Adverse Event Malfunction Summary report: N

BD Q-SYTE CLOSED LUER ACCESS DEVICE

MDR report key: 21930692 · Received April 29, 2025

Report

Report Number
1710034-2025-00703
Event Type
Malfunction
Date Received
April 29, 2025
Date of Event
April 7, 2025
Report Date
May 7, 2025
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FPA
UDI-DI
00382903851003
PMA / PMN Number
K013621
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 385100 AND LOT NUMBER 2182892. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD Q-SYTE CLOSED LUER ACCESS DEVICE SEPTUM DAMAGED/DISCOLORED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE PATIENT WAS RECEIVING TREATMENT IN OUR HOSPITAL. WHEN USING THE INFUSION CONNECTOR, THE NURSE FOUND THAT THE CONNECTOR HAD CHANGED COLOR AND WAS NOT SEALED. THE NURSE IMMEDIATELY REPLACED IT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1429239 BD Q-SYTE CLOSED LUER ACCESS DEVICE SET, ADMINISTRATION, INTRAVASCULAR FPA BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 2182892 00382903851003

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown