PERMANENT CAUTERY SPATULA INSTRUMENT
Report
- Report Number
- 2955842-2011-00256
- Event Type
- Malfunction
- Date Received
- August 5, 2011
- Date of Event
- July 8, 2011
- Report Date
- July 8, 2011
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K002489
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT ENGINEERING FOUND THE INSTRUMENT TO BE RETURNED WITH A DARK RED AND BROWN OBJECT APPROXIMATELY .095 INCHES BY .095 INCHES IN SIZE AND SEPARATED FROM THE INSTRUMENT. VISUAL INSPECTION OF THE OBJECT UNDER MAGNIFICATION CONCLUDED THAT THE OBJECT IS NOT PART OF THE INSTRUMENT, HOWEVER, IT IS A CHARRED PIECE OF BIO DEBRIS. ENGINEERING ALSO OBSERVED THE SPATULA EXHIBITS MELTING AND THERMAL DEFORMATION AT THE DISTAL END AND THE SURFACE HAS A DARK DISCOLORATION. THE CERAMIC SLEEVE WAS FOUND TO BE BROKEN AT THE DISTAL END, EXPOSING A SECTION OF THE SPATULA SHANK. BROKEN PIECES WERE NOT RETURNED WITH THE INSTRUMENT. THE PROXIMAL CLEVIS WAS FOUND TO BE CRACKED BETWEEN THE EARS, WITH THE CRACKING OBSERVED CONSISTENT WITH OVERLOADING AT THE TIP. ENGINEERING CONCLUDED THE DAMAGE FOUND WAS LIKELY DUE TO MISHANDLING AND OR MISUSE OF THE INSTRUMENT. PER THE CASE NOTES ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT DURING THE PROCEDURE.
IT WAS REPORTED THAT DURING A DA VINCI S PULMONARY WEDGE RESECTION PROCEDURE, WHILE USING THE PERMANENT CAUTERY SPATULA INSTRUMENT, THE SITE NOTICED A SMALL BLACK OBJECT INSIDE THE PATIENT. THE SURGEON REMOVED THE OBJECT AND BELIEVED IT TO BE PART OF THE PERMANENT CAUTERY SPATULA INSTRUMENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERMANENT CAUTERY SPATULA INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420184-04 | S10091008 380 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI S SYSTEM INSTRUMENTS, ACCESSORIES, & ESU |