FDA Adverse Event Malfunction Summary report: N

VASOVIEW HEMOPRO EVH SYSTEM

MDR report key: 2193017 · Received July 21, 2011

Report

Report Number
2242352-2011-00934
Event Type
Malfunction
Date Received
July 21, 2011
Date of Event
May 31, 2011
Report Date
May 31, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K052274
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE JAWS WERE SOMEWHAT BURNT. THERE WAS SOME EVIDENCE OF BLOOD. A PRE-CAUTERY TEST WAS PERFORMED ON THE DEVICE WITH A REFERENCE POWER SUPPLY AND EXTENSION CABLE. THE DEVICE PASSED THE PRE-CAUTERY TEST, IT PRODUCED STEAM AND HEAT DURING SEVERAL ACTIVATIONS AND SHUT OFF WHEN THE TOGGLE WAS RELEASED. BASED UPON THE FUNCTIONAL TEST, THE REPORTED FAILURE "JAWS MIS ALIGNED AND POOR SEALING" COULD NOT BE CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VESSEL HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO JAWS MIS-ALIGNED RESULTING IN POOR VESSEL SEALING. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VASOVIEW HEMOPRO EVH SYSTEM ENDOSCOPIC VESSEL HARVESTING SYSTEM GEI MAQUET CARDIOVASCULAR, LLC VH-3000-W 25026168

Patients

Seq Age Sex Outcome Treatment
1 NA