FDA Adverse Event Death Summary report: N

LUMEX

MDR report key: 2193005 · Received July 25, 2011

Report

Report Number
2428983-2011-00002
Event Type
Death
Date Received
July 25, 2011
Date of Event
June 23, 2011
Report Date
July 22, 2011
Manufacturer
APEX HEALTH CARE MFG INC.
Product Code
FNG
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

GF HEALTH PRODUCTS, INC. (DEVICE IMPORTER/DISTRIBUTOR) SENT A QUALITY MANAGER TO THE PT'S HOME ON TUESDAY (B)(6) 2011 TO INSPECT THE LF 1030 HYDRAULIC LIFT, TAKE DIGITAL PHOTOGRAPHS, VIDEO, AND INTERVIEW THE CAREGIVER AND PT'S DAUGHTER. THE ATTACHED REPORT REFLECTS NO MANUFACTURING DEFECT. THE PT SHOULD NOT HAVE BEEN LEFT UNATTENDED IN THE LIFT. THE MAINTENANCE SCHEDULE, SAFETY INSTRUCTIONS AND WARNINGS OUTLINED IN THE OWNER'S INSTRUCTION MANUAL HAD NOT BEEN FOLLOWED. THE INITIAL REPORTER (DEALER) HAS SIGNED ACKNOWLEDGEMENT OF RECEIPT. THE LIFT IS SCHEDULED TO BE PICKED UP ON TUESDAY, (B)(6) 2011 BY A GF HEALTH PRODUCTS, INC. REPRESENTATIVE AND SENT TO (B)(6) FOR AN INDEPENDENT THIRD PARTY EVAL.

Description of Event or Problem · 1

ON (B)(6) 2011 AT APPROXIMATELY 9 AM, THE CAREGIVER STATED SHE HAD THE PT IN AN UPRIGHT POSITION IN THE SLING AND ON THE LIFT. THE CAREGIVER WAS ATTEMPTING TO MOVE THE PT FROM THE BED TO THE DINING CHAIR. AFTER THE PT WAS ON THE LIFT, SHE STATED SHE PUT THE REAR CASTER BRAKES ON, AND LEFT THE PT (UNATTENDED) TO GET THE DINING CHAIR LOCATED BY THE BEDROOM DOOR. WHEN THE CAREGIVER TURNED SHE NOTICED THE LIFT WAS MOVING TOWARDS THE DOOR. SHE RUSHED TO THE FRONT OF THE LIFT AND TRIED TO STOP IT. SHE ATTEMPTED TO GRAB THE HANDLES. THE CAREGIVER WAS UNABLE TO STOP THE LIFT. SHE STATED THE LIFT HIT THE DOOR AND THE PT FELL ON THE FLOOR HITTING HIS HEAD BY THE DOOR FRAME. THEN THE LIFT UPPER ASSEMBLY FELL OVER ON THE PT. AFTER THE INCIDENT, THE CAREGIVER CARRIED THE PT FROM THE FLOOR TO THE BED AND TELEPHONED THE PT'S DAUGHTER. THE PT WAS LATER TRANSPORTED TO THE HOSPITAL AND DIED AT APPROXIMATELY 7 PM FROM INTERNAL BLEEDING IN THE PLURAL CAVITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LUMEX HYDRAULIC PATIENT LIFT FNG APEX HEALTH CARE MFG INC. LF 1030 31422

Patients

Seq Age Sex Outcome Treatment
1 95 YR Death