HS III PROXIMAL SEAL SYSTEM 3.8MM
Report
- Report Number
- 2242352-2011-00983
- Event Type
- Malfunction
- Date Received
- July 21, 2011
- Date of Event
- June 25, 2011
- Report Date
- June 28, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K080169
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE DELIVERY DEVICE WAS RETURNED SEPARATE FROM THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE INSIDE THE BODY OF THE LOADING DEVICE. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE PARTIALLY DEPRESSED ON THE DELIVERY DEVICE. THERE WAS SLIGHT EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL FAILED TO LOAD" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).
THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III SEAL FAILED TO LOAD, WHEN THEY WENT TO DEPLOY FROM THE HOLDER, IT GOT STUCK IN THE LOADER INSTEAD OF WHERE IT WAS SUPPOSED TO BE. A NEW KIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HS III PROXIMAL SEAL SYSTEM 3.8MM | CLAMPLESS BEATING HEART | GEI | MAQUET CARDIOVASCULAR, LLC | HSK-3038 | 25031748 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |