FDA Adverse Event Malfunction Summary report: N

VASOVIEW HEMOPRO EVH SYSTEM

MDR report key: 2192933 · Received July 21, 2011

Report

Report Number
2242352-2011-01003
Event Type
Malfunction
Date Received
July 21, 2011
Report Date
June 29, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K052274
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WILL REPORTEDLY NOT BE RETURNED TO MAQUET CARDIAC SURGERY FOR INVESTIGATION. THE LOT NUMBER COULD NOT BE OBTAINED, SO A LOT HISTORY RECORD REVIEW COULD NOT BE PERFORMED. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO UNIT WOULD NOT SHUT OFF. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS. THE PRODUCT IS NOT RETURNING. THE LOT NUMBER AND EVENT DATE ARE UNK, BUT THE INCIDENT OCCURRED SOME TIME IN (B)(6) 2011.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VASOVIEW HEMOPRO EVH SYSTEM ENDOSCOPIC VESSEL HARVESTING SYSTEM GEI MAQUET CARDIOVASCULAR, LLC VH-3000 NI

Patients

Seq Age Sex Outcome Treatment
1 NA