FDA Adverse Event Malfunction Summary report: N

VASOVIEW HEMOPRO 2 EVH SYSTEM

MDR report key: 2192931 · Received July 21, 2011

Report

Report Number
2242352-2011-00995
Event Type
Malfunction
Date Received
July 21, 2011
Date of Event
June 29, 2011
Report Date
June 29, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K101274
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE HEATER TIP WAS DETACHED. THE TIP OF THE COLD JAW BOOT HAD DETACHED. THE JAWS WERE SOMEWHAT BURNT. THERE WAS SOME EVIDENCE OF BLOOD. BASED UPON THE VISUAL OBSERVATIONS, THE REPORTED COMPLAINT FOR "HEATER SEPARATING" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NON-CONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VESSEL HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO 2 METAL PART WAS SEPARATING FROM THE JAWS. THE REPORTED UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VASOVIEW HEMOPRO 2 EVH SYSTEM ENDOSCOPIC VESSEL HARVESTING SYSTEM GEI MAQUET CARDIOVASCULAR, LLC VH-4000 25036291

Patients

Seq Age Sex Outcome Treatment
1 NA