WARMTOUCH COMPACT WARMING UNIT
MDR report key: 2192917
·
Received July 22, 2011
Report
- Report Number
- 2936999-2011-00494
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 22, 2011
- Manufacturer
- COVIDIEN/FORMERLY TYCO HEALTHCARE
- Product Code
- DWJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE VISUAL ALARM WORKS, THE UNIT IS NOT DESIGNED TO HAVE AN AUDIBLE ALARM. THE THERMISTOR AND POWER CORD ARE REPLACED. THE MFR DATE IS UNK AS THIS IS AN OLD SPARTAN UNIT. IT WOULD HAVE BEEN MANUFACTURED PRIOR TO 2002. THE WT-5800 WARMTOUCH IS NOT DISTRIBUTED IN THE U.S.; HOWEVER, THE WT-5300 WARMTOUCH IS OF ESSENTIALLY IDENTICAL DESIGN AND WAS DISTRIBUTED IN THE U.S. THE 510K NUMBER FOR THE WT-5300 WARMTOUCH IS (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED TO COVIDIEN THAT THE UNIT WILL HEAT AND WON'T STOP HEATING. WHEN THE TEMPERATURE IS REACHED, THE UNIT WILL NOT HOLD THE TEMPERATURE AND GOES ON HEATING. THE UNIT WON'T COOL BACK TO INPUT VALUES AND THE ALARM IS NOT GIVEN. THERE WAS NO PT HARM, NO PT WAS INVOLVED. THIS WAS FOUND WHEN CHECKING THE UNIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WARMTOUCH COMPACT WARMING UNIT | CONVECTIVE AIR WARMER | DWJ | COVIDIEN/FORMERLY TYCO HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |