FDA Adverse Event Malfunction Summary report: N

WARMTOUCH COMPACT WARMING UNIT

MDR report key: 2192917 · Received July 22, 2011

Report

Report Number
2936999-2011-00494
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 1, 2011
Report Date
June 22, 2011
Manufacturer
COVIDIEN/FORMERLY TYCO HEALTHCARE
Product Code
DWJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE VISUAL ALARM WORKS, THE UNIT IS NOT DESIGNED TO HAVE AN AUDIBLE ALARM. THE THERMISTOR AND POWER CORD ARE REPLACED. THE MFR DATE IS UNK AS THIS IS AN OLD SPARTAN UNIT. IT WOULD HAVE BEEN MANUFACTURED PRIOR TO 2002. THE WT-5800 WARMTOUCH IS NOT DISTRIBUTED IN THE U.S.; HOWEVER, THE WT-5300 WARMTOUCH IS OF ESSENTIALLY IDENTICAL DESIGN AND WAS DISTRIBUTED IN THE U.S. THE 510K NUMBER FOR THE WT-5300 WARMTOUCH IS (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED TO COVIDIEN THAT THE UNIT WILL HEAT AND WON'T STOP HEATING. WHEN THE TEMPERATURE IS REACHED, THE UNIT WILL NOT HOLD THE TEMPERATURE AND GOES ON HEATING. THE UNIT WON'T COOL BACK TO INPUT VALUES AND THE ALARM IS NOT GIVEN. THERE WAS NO PT HARM, NO PT WAS INVOLVED. THIS WAS FOUND WHEN CHECKING THE UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WARMTOUCH COMPACT WARMING UNIT CONVECTIVE AIR WARMER DWJ COVIDIEN/FORMERLY TYCO HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1