PEN NDL 32G 4MM 90 CT MAIL ORDER ONLY
Report
- Report Number
- 3023359743-2025-00325
- Event Type
- Malfunction
- Date Received
- April 29, 2025
- Report Date
- May 8, 2025
- Manufacturer
- EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND
- Product Code
- FMI
- PMA / PMN Number
- K213478
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ADDITIONAL INFORMATION WAS ADDED TO: B4, G6, H2, H3, H11. CORRECTION TO: H6 (COMPONENT CODE, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS) INVESTIGATION SUMMARY: SAMPLES WERE RECEIVED AND AN INVESTIGATION WAS PERFORMED. THIS IS THE 1ST COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE INDICATED ISSUE AS BENT NON PATIENT END CANNULA AND BASED ON TREND ANALYSIS NO FURTHER ACTION IS REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED.
CONSUMER REPORTED PEN NEEDLE CLOG WHEN PRIMING HIS PEN NEEDLES. LOT: 3241370, 4051715, 4044123. CATALOG: 320883. DATE OF EVENT: UNKNOWN. SAMPLES: YES CL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1596455 | PEN NDL 32G 4MM 90 CT MAIL ORDER ONLY | NEEDLE, HYPODERMIC, SINGLE LUMEN | FMI | EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND | 320883 | 4044123 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |