FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2192835
·
Received July 25, 2011
Report
- Report Number
- 1720753-2011-20390
- Event Type
- Malfunction
- Date Received
- July 25, 2011
- Date of Event
- July 18, 2011
- Report Date
- July 25, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE SVC REP REPLACED THE GENERATOR INTERFACE PRINTED CIRCUIT BOARD, AND THE NUMBER THREE POWER SUPPLY. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SVC.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYS WOULD EXHIBIT A POWER FAILURE. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |