FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2192666 · Received August 5, 2011

Report

Report Number
6000001-2011-16752
Event Type
Malfunction
Date Received
August 5, 2011
Date of Event
July 14, 2011
Report Date
July 20, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, BUT THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 807:05 WAS CONFIRMED DURING PRODUCT EVALUATION. THE ASSIGNABLE CAUSE FOR THE CONDITION WAS DETERMINED TO BE A FAULTY PUMP HEAD MODULE (PHM). THE PHM WAS REPLACED TO CORRECT THIS CONDITION. ADDITIONAL INFORMATION: THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90.

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE OF 807:05 THAT OCCURED FIVE TIMES. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, THERE WAS NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION REPORTED RELATED TO THE DEVICE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1