COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-16750
- Event Type
- Malfunction
- Date Received
- August 5, 2011
- Date of Event
- July 14, 2011
- Report Date
- July 18, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 550 WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THE REPORTED PROBLEM WAS DETERMINED TO BE DEFECTIVE UIM PCB (USER INTERFACE MODULE PRINTED CIRCUIT BOARD). APPROPRIATE ACTION WAS TAKEN TO CORRECT THE REPORTED PROBLEM BY REPLACING THE UIM PCB. THE DEVICE HAS NOT BEEN PREVIOUSLY SENT INTO SERVICE FOR THE REPORTED CONDITION OF "550 ERROR CODE." THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 6.13.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006.
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, BUT THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A 550 ERROR CODE. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, THERE WAS NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION REPORTED RELATED TO THE DEVICE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |