FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2192657 · Received August 5, 2011

Report

Report Number
6000001-2011-16743
Event Type
Malfunction
Date Received
August 5, 2011
Date of Event
July 12, 2011
Report Date
July 13, 2011
Manufacturer
BAXTER HEALTHCARE-SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED AND IS AVAILABLE FOR EVALUATION. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH A BLANK MAIN DISPLAY WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS DETERMINED TO BE A FAULTY USER INTERFACE MODULE (UIM) PRINTED CIRCUIT BOARD (PCB). THE UIM PCB WAS REPLACED TO CORRECT THIS CONDITION. A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT A COLLEAGUE INFUSION PUMP IN WHICH THERE WAS A BLANK MAIN DISPLAY. THIS CONDITION HAS THE POTENTIAL TO CAUSE AN INTERRUPTION OF DELIVERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE-SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1