COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-16742
- Event Type
- Malfunction
- Date Received
- August 5, 2011
- Date of Event
- August 1, 2011
- Report Date
- August 1, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Removal / Correction Number
- 6000001-08/04/10-002-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY WAS CONFIRMED AND DUPLICATED BY BAXTER SERVICE PERSONNEL. THE ROOT CAUSE OF THIS CONDITION WAS DETERMINED TO BE DEPLETED MAIN BATTERIES AS A RESULT OF USER ERROR. THE MAIN BATTERIES AND HARNESS WERE REPLACED TO CORRECT THIS CONDITION. ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY. IT IS UNKNOWN WHEN, OR IN WHICH CARE AREA, THIS EVENT OCCURRED. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT DELIVERY. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO INFORMATION REGARDING PATIENT INVOLVEMENT, PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER REMEDIATED INTERFACE MODULE SOFTWARE VERSION IS 5.09.90 - COLLEAGUE 2006.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |