FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2192617 · Received August 5, 2011

Report

Report Number
1423500-2011-10315
Event Type
Malfunction
Date Received
August 5, 2011
Date of Event
July 4, 2011
Report Date
July 12, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR CONNECTION ISSUE WAS CONFIRMED IN THE LAB. THE RESULTS OF A SAMPLE EVALUATION REVEALED THAT THE PULL RING CAP WAS MISSING FROM THE DRAIN LINE. THE ASSIGNABLE CAUSE WAS DETERMINED TO BE A MANUFACTURING ISSUE. THE DEVICE WAS DESTROYED. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN REQUESTED FOR EVALUATION. A BATCH REVIEW WILL BE PERFORMED FOR THE LOT NUMBER PROVIDED. SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

A NURSE REPORTED TO BAXTER THAT A PATIENT LINE HAD A 'LOOSE COVERING'. THERE WAS NO PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.