COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-16747
- Event Type
- Malfunction
- Date Received
- August 5, 2011
- Date of Event
- July 11, 2011
- Report Date
- July 14, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K053696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP THAT EXPERIENCED A DAMAGED BATTERY ALARM WAS CONFIRMED BY BAXTER PERSONNEL DURING PRODUCT EVALUATION. THE ASSIGNABLE CAUSE FOR THE REPORTED PROBLEM WAS DETERMINED TO BE DEPLETED BATTERIES RESULTING FROM USER ERROR. THIS CONDITION WAS RESOLVED BY REPLACING THE MAIN BATTERIES AND BATTERY HARNESS. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90. A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THERE HAVE BEEN PREVIOUS REPORTED PROBLEMS WITH THIS DEVICE THAT ARE THE SAME AS OR SIMILAR TO THE REPORTED CONDITION.
(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT A COLLEAGUE INFUSION PUMP IN WHICH A DAMAGED BATTERY ALARM OCCURRED. THIS CONDITION HAS THE POTENTIAL TO CAUSE AN INTERRUPTION OF DELIVERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |