FDA Adverse Event
Malfunction
Summary report: N
1.8MM DRILL BIT WITH DEPTH MARK/QC/110MM
MDR report key: 2192597
·
Received July 1, 2011
Report
- Report Number
- 8030965-2011-00302
- Event Type
- Malfunction
- Date Received
- July 1, 2011
- Report Date
- June 7, 2011
- Manufacturer
- SYNTHES GMBH
- Product Code
- HTW
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
SUBJECT DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED. INVESTIGATION CO-ORDINATED BY SYNTHES (B)(4). REPORT RECEIVED INDICATES THAT THE TIP OF THE DRILL BIT PARTIALLY BROKE OFF. THE MICROSCOPIC VIEW OF THE BROKEN SURFACE DOES NOT SHOW ANY ANOMALIES OF THE MATERIAL QUALITY. THE AVAILABLE DIMENSIONS WERE CHECKED AND FOUND TO BE WITHIN SPECIFICATION. FURTHER INVESTIGATION REGARDING THE MFG AND RAW MATERIAL DOCUMENTS SHOWS CONFORMITY TO THE SPECIFICATION. NO PRODUCT OR MATERIAL FAULT COULD BE DETECTED.
Description of Event or Problem · 1
THE 1.8 MM DRILL BIT BROKE WHILE DRILLING IN BONE. BROKEN BIT WAS LEFT IN PT'S BONE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 1.8MM DRILL BIT WITH DEPTH MARK/QC/110MM | QUICK COUPLING DRILL BIT | HTW | SYNTHES GMBH | NA | 2658836 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |