FDA Adverse Event Malfunction Summary report: N

1.8MM DRILL BIT WITH DEPTH MARK/QC/110MM

MDR report key: 2192597 · Received July 1, 2011

Report

Report Number
8030965-2011-00302
Event Type
Malfunction
Date Received
July 1, 2011
Report Date
June 7, 2011
Manufacturer
SYNTHES GMBH
Product Code
HTW
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SUBJECT DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED. INVESTIGATION CO-ORDINATED BY SYNTHES (B)(4). REPORT RECEIVED INDICATES THAT THE TIP OF THE DRILL BIT PARTIALLY BROKE OFF. THE MICROSCOPIC VIEW OF THE BROKEN SURFACE DOES NOT SHOW ANY ANOMALIES OF THE MATERIAL QUALITY. THE AVAILABLE DIMENSIONS WERE CHECKED AND FOUND TO BE WITHIN SPECIFICATION. FURTHER INVESTIGATION REGARDING THE MFG AND RAW MATERIAL DOCUMENTS SHOWS CONFORMITY TO THE SPECIFICATION. NO PRODUCT OR MATERIAL FAULT COULD BE DETECTED.

Description of Event or Problem · 1

THE 1.8 MM DRILL BIT BROKE WHILE DRILLING IN BONE. BROKEN BIT WAS LEFT IN PT'S BONE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 1.8MM DRILL BIT WITH DEPTH MARK/QC/110MM QUICK COUPLING DRILL BIT HTW SYNTHES GMBH NA 2658836

Patients

Seq Age Sex Outcome Treatment
1 NI