FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2192576 · Received August 5, 2011

Report

Report Number
1423500-2011-10313
Event Type
Malfunction
Date Received
August 5, 2011
Date of Event
July 16, 2011
Report Date
July 16, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER. THEREFORE, THE DEVICE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE SYSTEM ERROR 2240 (AIR IN LINE) WAS CONFIRMED, AND THE ROOT CAUSE WAS DETERMINED TO BE USE ERROR- PATIENT DISCONNECTED FROM THE SET. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER TECHNICAL SERVICES (BTS) REGARDING ASSISTANCE WITH A SYSTEM ERROR 2240 ALARM DURING DRAIN 4 OF 4 ON THE HOMECHOICE MACHINE (HC). THE NURSE REPORTED THAT THE HOME PATIENT (HP) DISCONNECTED AND FORGOT TO PRESS STOP, AND THEN RECONNECTED TO THE HC BEFORE THE ALARM OCCURRED. THE TECHNICAL SERVICE REPRESENTATIVE(TSR) ASSISTED THE NURSE TO CYCLE POWER TO GET TO "PRESS GO TO START." THERE WAS PATIENT INVOLVEMENT; HOWEVER, THERE WAS NO INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 38 YR HOMECHOICE