FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 2192572 · Received June 30, 2011

Report

Report Number
1824206-2011-03522
Event Type
Malfunction
Date Received
June 30, 2011
Date of Event
June 17, 2011
Report Date
June 17, 2011
Manufacturer
HILL-ROM INC.
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE FOOT END BRAKE LINKAGE WAS BROKEN. HE REPLACED THE BRAKE LINKAGE TO REPAIR THE STRETCHER.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE BRAKE WILL SET AND HOLD ON THE HEAD END OF THE STRETCHER BUT NOT ON THE FOOT END.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILL-ROM INC. 8050

Patients

Seq Age Sex Outcome Treatment
1