FDA Adverse Event Malfunction Summary report: N

INTERMATE

MDR report key: 2192531 · Received August 5, 2011

Report

Report Number
6000001-2011-16732
Event Type
Malfunction
Date Received
August 5, 2011
Date of Event
July 14, 2011
Report Date
July 15, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PER REVIEW OF THE BATCH RECORDS, NO NONCONFORMANCE REPORT WAS DOCUMENTED FOR THIS LOT. ALL RELEASE CRITERIA WERE MET FOR THE BUILD OF THE LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. THE REPORTER INITIALLY REPORTED THAT THE SAMPLE IS AVAILABLE. HOWEVER, THE SAMPLE WAS NOT RETURNED TO BAXTER FOR EVALUATION. THEREFORE, THE REPORTED CONDITION COULD NOT BE CONFIRMED, AND THE ROOT CAUSE COULD NOT BE DETERMINED. SHOULD THE DEVICE AND/OR ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS AVAILABLE FOR EVALUATION PER THE CUSTOMER; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INTERMATE LEAKED FROM THE BLUE WINGED LUER LOCK CAP. THIS CONDITION OCCURRED BEFORE PATIENT USE. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. THE SAMPLE IS AVAILABLE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTERMATE PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE 11B062

Patients

Seq Age Sex Outcome Treatment
1