FDA Adverse Event
Malfunction
Summary report: N
I-STAT 1 ANALYZER
MDR report key: 2192526
·
Received June 30, 2011
Report
- Report Number
- 2245578-2011-00102
- Event Type
- Malfunction
- Date Received
- June 30, 2011
- Date of Event
- June 6, 2011
- Report Date
- June 30, 2011
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- JJE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED BY A DISTRIBUTOR WHO REPORTED THAT AN I-STAT 1 ANALYZER SHOWED OUT-OF-SPEC STANDBY CURRENT. THE ANALYZER WAS RETURNED TO APOC AND, DURING ROUTINE FAILURE ANALYSIS ON (B)(4) 2011, A BURNT CAPACITOR WAS FOUND. BASED ON THE INFORMATION AVAILABLE AT THE TIME, THERE WAS NO INJURY ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT 1 ANALYZER | I-STAT 1 ANALYZER | JJE | ABBOTT POINT OF CARE INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |